Very low readings across the plate in ELISA experiment


Author:Elabscience
views:559

Possible causes
Solutions
Incorrect wavelength
Check filters/reader
Stop solution not added
Add stop solution to each well
Insufficient incubation timeIncrease incubation time
Incubation temperature vary wildly
Check the incubator
Incorrect dilutions of detection Ab/HRP-conjugate
Adhere to the manual
The kit has lost its activity
Get a new kit